Biogas Online Analyzer BGA-1000F
-
Brand:LKS (M) Sdn. Bhd.
Description
CONTENT
1. Brief
2. System flow
3. Complete set
4. Technical parameter
5. Installation
6. Operation
7. Maintenance
8. Transportation and storage
9. Quality assurance
10. Detail Internal Layout (Appendix 1)
Features
Analysis cabinet’s housing is stainless steel which is adaptable to corrosive environment. Besides, cabinet is sealed to prevent from corrosive gas. Intelligent
heater is built inside to ensure normal functioning of the system when inside temperature approaches 8℃. Analysis system can be programmed to realize auto‐drain and auto zero calibration function. Online infrared biogas analyzer is the core part of the system. Gas molecules Like CO2、CH4 which are composed of different types of atoms have absorption spectrum in infrared range. Absorption
intensity observes Lamber‐Beer’s Law. Our product is on the basis of it. In addition, electrochemical oxygen sensor measuring module can be installed internally so as H2S Sensor. One single analyzer can measure up to 4 gas components. Our equipment can be applied to detect the concentration of landfill gas, biogas and CO2、CH4、H2S、O2 produced in the biogas generating process.
It is applicable to various kinds of environment. Through analyzing parameter change, it helps to judge and alarm in industrial process. Meantime, 4‐20mA DC output and digital alarm output interface are available.
Working environment
• Power Supply : 230VAC, 50HZ to 60HZ
• Sample gas specification:‐
• sampling pressure : 3 ~ 20Kpa
• exhaust pressure : 0 ~ ‐10Kpa
• Operating temperature : 0 ~ 80C
• Ambient temperature : ‐10C ~ 55C
Flow Chart - Fig 1
Specification :
• Globe valve V1,V2
Type: Manual ball valve, stainless steel;
Function: Cut off sampling and exhaust pipeline when doing maintenance
• Throttling valve V5
Type: Ø6 stainless steel needle valve, cutting sleeve connector;
Function: adjust flow to a level of 0.7~1.2L/min
• Gas-water separator FIT1
Type: PTFE core, Filter size: 3um;
Function: Separate liquid water and filter impurities of 3um
• Ultra-filter FIT2
Type: Fibre core,Filter size: 1um;
Function: Filter impurities of 1um and above
• Solenoid valve V3,V4
Type: DC24V,normally closed;
Function: It is programmed to regularly evacuate water automatically.
• Heater JRQ
Type: AC220V,300W;
Function: Avoid pipeline freezing and plugging when the inside temperature is below 8C
Complete System
Equipment and accessories:
No Name
1) Biogas Analyzer
Model : BGA‐1000F
Qty : 01 unit
2) Manual Ball Valve
Model : 4‐way, Stainless steel
Qty : 02 unit
3) Reducer Union
Model : ½”/Ø6mm Stainless Steel
Qty : 02 unit
4) Sampling and Exhaust pipe
Model : ½”/Ø6mm PTFE
Qty : 01 meter
5) Insulation Pipe
Qty : 01 meter
6) Throttling valve
Model : Ø6mm cutting sleeve, Stainless Steel
Qty : 01 unit
7) Spare Part
7.1) Core
Model : Ø6mm
Qty : 04 units
7.2) Filter core of gas water
Model : PTFE
Qty : 01 unit
7.3) Hexagon Wrench
Model : M8
Qty : 01 unit
8) Documents
8.1) Manufacturer Certificate
Qty : 01 book
8.2) Instrument manual
Qty : 01 cert
Technical data
Gas Parameter: CH4 Methane
Range:0~100%(VOL)
Principle: NDIR
Accuracy: ±1%FS
Nonlinearity: ≤±1%FS
Zero drift: ≤± 1%FS / per week
Output: 4~20mA( maximum 750 Ω), DC/alarm output 2 point
Gas Parameter: CO2 Carbon Dioxide
Range:0~50%(VOL)
Principle: NDIR
Accuracy: ±1%FS
Nonlinearity: ≤±1%FS
Zero drift: ≤± 1%FS / per week
Output: 4~20mA(maximum 750 Ω),DC/ alarm output 2 point
Gas Parameter: H2S Hydrogen Sulphide
Range: 0~1%(VOL)
Principle: ECD
Accuracy: 3%FS
Zero drift:≤0.05%(VOL)/per week
Temp. drift:≤±0.025%(VOL)
nonlinearity: ≤±0.5% full range
Output: 4~20mA(maximum 750 Ω),DC/ alarm output 2 point
Gas Parameter: O2 Oxygen
Range:0~25%(VOL)
Principle:ECD
Accuracy: 3%FS
Zero drift:≤0.05%(VOL)/per week
Temp. drift:≤±0.025%(VOL)
Nonlinearity: ≤±0.5%full range
Output: 4~20mA(maximum 750 Ω),DC/ alarm output 2 point
Sampling gas pipe installation
• To avoid mechanical damage, it’s appropriate to lay the sampling pipe and exhaust pipe along the support frame or pipe chase.
• If it is difficult to recycle the exhaust to process pipeline, you can evacuate it in safe area.
• You can not apply in‐site air as zero gas which is severely polluted. Collect clean air with extra pipe line if it is so.
• If it is in open area, there is no need to connect extra pipeline to drain the water.
Electrical installation
• Panel
Power Consumption: 500Watt
Power Supply: 230VAC, 50‐60Hz
Power cord is led through power entrance to main switch QF up end, cable model:RVV 3*1.5mm2
• Signal cable
Apply cable model RVVP 8*0.75mm2 if it is 4~20mA signal output. If it is digital alarm signal, you should choose according to alarm component.
There are 4 alarm cables in total. Connecting terminal lies under the panel inside the cabinet. Refer to instrument manual for specific wiring.
Operation
To ensure the pipeline is without leakage, check carefully after all installation work is done. Make sure that grounding work is done and power supply meets particular requirements.
Power on
Switch MCB and enter in warm‐up interface, then system starts to work. Close the cabinet door and tighten it with hexagon wrench. It takes 10min’s warm‐up time to enter in measuring interface.
Sampling
First close V5 and then open V2, V1 in sequence, rotate V5 gently eventually till the flow is 1.0L/min. sampling gas will flow into analysis panel, it will show the gas components reading after a few seconds.
Maintenance
Instrument maintenance
Signal drift will occur if the analysis instrument is regularly used. It contains automatic zero calibration function. If you need calibrate manually, please refer
to concerning chapter.
Pipeline maintenance
To ensure reliable functioning of the system, you should check the pipeline working condition regularly to make sure that there is no leakage and plugging.
You can judge the trouble through below:
• If there is tiny flow or even no flow in the flow meter, please check whether the process pipeline pressure is normal first and check whether the pipeline is choked. Special notice: This trouble will also appear if exhaust pipe is choked.
• If there is a huge discrepancy between measurement result and experience. There may be leakage occurring.
Filter core maintenance
You should replace the filter material regularly according to actual pollution condition. Cut off sample gas valve and exhaust valve before replacing. To ensure tightness, lock cabinet door after replacement.
Drain
In cold weather condition, make sure there is no frost in drain
Shipment and storage
Gas analysis system is highly sophisticated equipment. You should pack with wooden case stuffed with vibration‐absorptive material so that the glass will not be pressured against .Ensure the cabinet is in upright state in the process of shipment, anti‐moisture measures should be taken to protect electrical unit.
Quality Guarantee
After‐sales service commitment :
• Calculated from the day when the products are inspected and accepted, warrantee period is one year.
• We will bear the freight cost and repair the products for free due to quality reasons if it is under warrantee.
• Engineer on site will be charge according to man day event the product is still under warranty.
• After warrantee period, we still assume repair responsibility, but cost and freight fee will be charged.